What scratch hardness tells you
Lenses, screens, coatings, polished metals and plastic housings are judged by how well they resist scratching. Indentation hardness measures resistance to a static load, but scratching adds sliding, and with it ploughing, pile-up and fracture. Scratch hardness measures that combined resistance directly, giving a number that tracks real scratch damage. The scratch testing overview explains the test, the instrument and the other scratch methods.
How the test works
- A low-load profile scan records the original surface along the scratch path.
- A conical diamond stylus of defined geometry is drawn across the surface at a constant normal load, at a set speed and length.
- A second low-load scan, an optical image or a 3D profilometer scan measures the residual scratch width.
- Scratch hardness is calculated from the load and the residual width, per ASTM G171.
Each sample is scratched several times under identical conditions to confirm reproducibility.
Comparing materials
Because the calculation uses only load and residual width, materials as different as aluminium and stainless steel can be ranked directly. In NANOVEA’s metal study on a PB1000, a 120° diamond cone with a 200 µm tip gave scratch widths of 220 µm on Cu110, 174 µm on Al6061 and 89 µm on SS304. That corresponds to scratch hardness of 0.52, 0.84 and 3.2 GPa.
Measuring the scratch properly
Scratch width is the key input, so how it is measured matters. Optical measurement is quick; low-load profile scans give the profile change directly on the instrument; 3D non-contact profilometry adds depth and pile-up and avoids judging edges by eye. For polymers with strong elastic recovery, the residual width must be measured after the stylus has passed, not under load.
Mechanical tester or tribometer
NANOVEA mechanical testers cover scratch hardness at nano and micro loads, suited to thin layers, polymers and small samples. For higher loads and testing at temperature, ASTM G171 scratch hardness also runs on NANOVEA tribometers, including inside the high-temperature oven.