Mechanical testing · Low temperature module

Low Temperature Nanoindentation & Scratch

The low temperature module cools the indentation contact in an enclosed Peltier-cooled environment, to -10 °C as standard and below -40 °C as a custom option, with tip and sample inside the same enclosure. Hardness, elastic modulus, creep and scratch resistance can then be measured at the sub-zero temperatures where polymers, elastomers and adhesives become stiffer or more brittle.

Low temperature enclosure for NANOVEA mechanical testers

At a glance

Temperature
Down to -10 °C as standard; below -40 °C as a custom configuration.
Cooling
Enclosed Peltier cooling system for stable, accurate temperature control.
Configuration
Indenter tip and sample both inside the enclosed, cooled environment.
Tests
Nanoindentation hardness and modulus, creep, nanoindentation DMA and scratch.
Materials
Polymers, elastomers, adhesives, coatings and food or biological materials used cold.
Platforms
PB1000 and CB500 mechanical testers.

Why test mechanical properties in the cold

Automotive seals and gaskets, aerospace polymers, outdoor coatings, adhesives, frozen food and cold-stored medical products all work below room temperature. Cooling stiffens polymers and elastomers, slows their relaxation and can make them brittle. Room-temperature hardness and modulus can therefore be misleading for anything used cold.

How the module works

The indentation contact is enclosed in a Peltier-cooled environment. The indenter tip and the sample sit together inside the enclosure, so both reach the set temperature before testing. The standard configuration reaches -10 °C; colder configurations, below -40 °C, are available as custom options.

Thermoelectric cooling suits nanoindentation: it has no compressor at the instrument to add vibration, and it is precisely controlled electronically, so the temperature stays stable during the long holds used for creep.

Keeping the measurement stable

As at high temperature, the main risk at low temperature is thermal drift: any temperature difference between tip and sample makes them expand or contract during the test, adding false displacement. Enclosing both together and letting the system equilibrate before testing keeps the depth signal reliable. Limiting exposure to humid room air also limits condensation or frost on the sample surface.

What to measure

The same tests as at room temperature can run in the cold: hardness and elastic modulus, creep and relaxation, nanoindentation DMA as materials approach their glass transition, and scratch tests for coatings that may crack when cold. For friction and wear below zero, see NANOVEA’s low temperature tribology module.

Tests you can run below ambient

Instruments that run this test

Every instrument is built to order. Send us your samples and we run your test in our laboratory first, then configure the instrument around the result.

Independent published research

NANOVEA instruments appear in 438 published studies on indentation, hardness and scratch, by researchers at universities and industrial laboratories worldwide.

All 438 studies →

NANOVEA materials testing laboratory

Need sub-zero mechanical data? Send us your samples.

Our laboratory runs low temperature indentation and scratch tests on the same mechanical testers we build, and reports results with the full temperature conditions and test parameters.

Your samples are tested by the people who design the instruments, so you get advice on the method as well as the numbers.

Indentation & scratch lab services

Frequently asked questions

How cold can the mechanical tester go?

The enclosed Peltier cooling system reaches -10 °C as standard. Lower temperatures, below -40 °C, are available as a custom configuration.

Why test hardness and modulus at low temperature?

Polymers, elastomers and adhesives stiffen and can become brittle as they cool toward their glass transition. A seal, adhesive or coating that is compliant and tough at room temperature can crack or lose grip in the cold, so its properties must be measured at service temperature.

Why is the tip inside the cooled enclosure?

If the tip were at room temperature and the sample cold, heat would flow through the contact during the measurement, causing thermal drift in the depth signal. Keeping tip and sample in the same enclosed environment brings them to the same temperature, which keeps depth, modulus and creep measurements reliable.

Why Peltier cooling?

Thermoelectric (Peltier) cooling has no compressor or coolant flow at the instrument, so it introduces little vibration and allows precise electronic control of temperature, both important for nanoindentation.

What about condensation and frost?

Moisture can condense or freeze on a cold sample and affect surface measurements. The enclosed environment limits exposure to humid room air, and samples are allowed to equilibrate inside it before testing.