Tribology testing · Liquid module

Lubricated & Liquid Tribology Testing

The liquid module runs friction and wear tests with the contact immersed in a liquid, or fed with lubricant drop by drop. It covers lubricant performance, wear in water, body fluids or process liquids, and the full lubrication regime map: a continuous Stribeck curve can be measured in a single run.

At a glance

Immersion
Liquid cups for rotational and linear tests keep the contact fully submerged.
Heating
A heating coil heats the liquid in the cup up to 150 °C.
Drop-by-drop
Metered lubricant feed for a thin film at high speed, usable even at 5000 rpm where immersion is impractical.
Stribeck curves
Step-less speed control sweeps from high to low speed in one run, giving a continuous Stribeck curve.
Modules
Rotational, linear and scratch; block-on-ring and ring-on-ring as custom configurations.
Measurements
Friction versus time and speed, Stribeck curves, wear rate from 3D profilometry.

What lubricated testing tells you

Most real tribological contacts are lubricated, whether by oil, grease, water, a process fluid or a body fluid. Lubricated tests answer how much the fluid reduces friction and wear, under which conditions its film breaks down, and how surfaces and fluids should be paired. The difference can be dramatic: in NANOVEA’s block-on-ring study, heavy mineral oil cut friction from ~0.8 to 0.09 and reduced wear by orders of magnitude.

Two ways to lubricate the contact

Immersion. The sample sits in a liquid cup, so the contact is fully submerged throughout the test. Cups are available for rotational and linear tests, and a heating coil brings the liquid up to 150 °C. Immersion suits lubricant ranking, wear in water or saline, implant materials in simulated body fluids, and Stribeck curves.

Drop-by-drop. A metered feed applies a constant amount of lubricant to the contact. It reproduces starved or thin-film lubrication and works at high speed, even 5000 rpm, where a bath would churn or fail to form a realistic film. NANOVEA’s piston wear study used drop-by-drop feed on the T2000’s linear reciprocating stage to simulate the piston skirt and cylinder liner contact.

Stribeck curves in one run

A Stribeck curve maps friction across the lubrication regimes as speed changes: boundary, mixed and hydrodynamic. Traditionally it is assembled from many separate constant-speed tests, each with its own run-in and set-up variation. With step-less speed control, a NANOVEA tribometer ramps speed continuously, for example exponentially from 15,000 down to 0.01 rpm. The complete curve comes from one run of about 10 minutes, on one track, under one set of conditions.

Dry versus wet

The same material can rank differently dry and wet. In NANOVEA’s sandpaper study, two products of the same grit, both sold for dry and wet use, swapped places: one abraded better dry, the other wet, and in each case the condition with higher friction gave the higher abrasion rate. Testing in the actual service condition avoids that kind of surprise.

Measuring wear after lubricated tests

Lubricated wear tracks can be shallow and smooth, which makes mass loss unreliable. After cleaning, the track is measured by chromatic confocal 3D non-contact profilometry, which resolves small worn volumes and the fine morphology of polished or mildly worn surfaces. The specific wear rate follows from K = V / (F × s).

Tests you can run in liquid

Instruments that run this test

Every instrument is built to order. Send us your samples and we run your test in our laboratory first, then configure the instrument around the result.

Published results

Measured in our own laboratory, on the instruments we design and build, and published with full test parameters.

All tribology application notes →

Independent published research

NANOVEA instruments appear in 45 published studies on lubricated and liquid tribology, by researchers at universities and industrial laboratories worldwide.

All 45 studies →

NANOVEA materials testing laboratory

Comparing lubricants or fluids? Send us your samples.

Our laboratory runs lubricated friction and wear tests, Stribeck curves and immersed-wear tests on the same instruments we build, and reports results with the full test parameters.

Your samples are tested by the people who design the instruments, so you get advice on the method as well as the numbers.

Friction & wear lab services

Frequently asked questions

What is a Stribeck curve?

A plot of friction coefficient against a lubrication parameter, usually viscosity × speed / load. Read from low to high values it shows boundary lubrication (high friction, asperity contact), mixed lubrication (friction falling as a film builds) and hydrodynamic lubrication (a full film, with friction rising slowly from viscous drag). It characterises a lubricant and surface pair across operating conditions.

How is a continuous Stribeck curve measured?

Instead of running separate tests at fixed speeds, the tribometer sweeps speed continuously with a step-less exponential ramp, for example from 15,000 down to 0.01 rpm. Friction is recorded throughout, so one run of about 10 minutes yields the complete curve.

When should drop-by-drop lubrication be used instead of immersion?

When the contact moves too fast for immersion to form a realistic thin film, or when the application is lubricated by a limited supply rather than a bath: piston skirts against cylinder liners, for example. The drop-by-drop feed applies a constant amount of lubricant to the contact and works even at 5000 rpm.

How hot can the liquid be?

A heating coil heats the liquid in the cup up to 150 °C, so lubricants and fluids can be tested at operating temperature.

What liquids can be tested?

Oils and greases, fuels, water and aqueous solutions, biological fluids and simulants for implants, and process or cleaning liquids. For corrosive electrolytes where electrochemistry matters, use the tribocorrosion module.

Which test configurations work with the liquid module?

Rotational (pin-on-disk), linear reciprocating and scratch tests, with block-on-ring and ring-on-ring available as custom configurations.