Application Note | Sensor Inspection
Quartz Crystal Microbalance Surface Flatness & Finish
Non-Contact Flatness, Finish and Defect Inspection of a Nanogram-Sensitive Sensor

Research & Experimental Testing
Duanjie Li
Visual Design & Editorial
Andrew Shore
Introduction
The Quartz Crystal Microbalance (QCM) is a key component used in a variety of research and industrial instruments to detect and monitor the small variation of mass, adsorption, density and corrosion, etc. As an extremely sensitive mass sensor capable of measuring mass change in the nanogram range, the manufacturing of QCM requires stringent quality control to ensure its precision. Any surface defect on the QCM plate may cause significant measurement inaccuracy and errors.
Importance of 3D Non-Contact Profilometry for Quality Control of QCM
Reliable quality control heavily relies upon accurate, quantifiable and reproducible surface inspection. Flatness and finish of the QCM surface are vital to its accuracy and both measurements in 3D guarantee proper manufacture processing and control measures. Unlike a touch probe technique, the NANOVEA Profilometer performs a 3D non-contact surface measurement of the sample. This eliminates the risk of creating micro scratches on the QCM surface that may cause inaccuracy or errors in the mass measurement.
ℹ️ Learn more about surface profiling lab services.
Measurement Objective
In this application, the NANOVEA ST400 is used to measure the surface flatness and finish of a QCM sample. We showcase the capacity of the NANOVEA non-contact profilometer in providing 3D profile measurement on delicate samples with high precision.

Results and Discussion
An array of 3 × 3 images is automatically taken and stitched by the NANOVEA ST400, as shown in Fig. 2, for a larger microscopic observation of the QCM sample. The area of interest is then chosen for 3D scan.

The false color view and the 3D view of the QCM sample surface are shown in Fig. 3, allowing users to observe the QCM sample from different angles and magnifications. Flatness variation can be seen clearly by color and height changes at the top of the QCM sample. The center of the QCM sample is slightly elevated compared to the edge. The flatness is 246 µm according to the ISO 12781 standard, as calculated by the NANOVEA analysis software. The QCM is an extremely sensitive and accurate device capable of detecting mass change down to 0.1 nanogram. The flatness and uniformity of the QCM play an important role in ensuring its quality and accuracy.

The 3D non-contact surface scan provides a useful tool to determine the surface finish quality with high precision. Fig. 4 shows the QCM sample surface after form removal. The finish values are calculated and summarized in Table 1. Fig. 5 presents the surface morphology of the QCM sample at a high magnification. Depending on the applications of the QCM, such as monitoring thin film deposition and cell growth, the QCM requires different surface finishes. The capacity of high resolution scan enables users to measure the detailed morphology and tiny surface features. Moreover, it can pick up any undesired defects including scratches or cracks on the sample surface.

| ISO 25178 height parameter | Value |
|---|---|
| Sa | 0.654 µm |
| Sq | 0.868 µm |
| Sp | 8.29 µm |
| Sv | 8.40 µm |
| Sz | 16.7 µm |
| Ssk | −0.169 |
| Sku | 9.50 |
Table 1: Summary of finish values.

Fig. 6 displays the light intensity map of the QCM sample surface. The metal electrode shows a much higher intensity compared to the quartz substrate. This function provides users additional information that may be used to determine the different materials or phases on the sample surface.

Conclusion
In this application, we have showcased that the NANOVEA ST400 3D Non-Contact Profilometer performs surface inspection on a QCM sample with high accuracy. The automatic microscope image array enables observation of a larger surface under the microscope. The 3D non-contact surface scan is a very useful quality control tool to determine whether the manufactured QCMs fall in line with specifications including finish and flatness.
The data shown here represents only a portion of the calculations available in the analysis software. NANOVEA Profilometers measure virtually any surface in fields including Semiconductor, Microelectronics, Solar, Fiber Optics, Automotive, Aerospace, Metallurgy, Machining, Coatings, Pharmaceutical, Biomedical, Environmental and many others.
Frequently Asked Questions About QCM Surface Inspection
Why does QCM surface quality matter?
A quartz crystal microbalance detects mass changes in the nanogram range, down to 0.1 nanogram; any surface defect on the plate can cause significant measurement errors, so manufacturing needs stringent quality control.
Why inspect a QCM without contact?
A touch probe risks micro scratches on the QCM surface that could themselves cause inaccuracy in the mass measurement; an optical profilometer measures flatness and finish without touching.
How flat was the QCM?
The center was slightly elevated compared to the edge, with a flatness of 246 µm according to ISO 12781.
What surface finish did it have?
After form removal, Sa 0.654 µm and Sq 0.868 µm (ISO 25178). Different QCM applications, such as thin film deposition monitoring and cell growth, require different finishes.
What does the intensity map show?
The metal electrode reflects much more light than the quartz substrate, so the intensity map helps distinguish different materials or phases on the surface.
How is a larger area located?
The ST400 automatically takes and stitches a 3 × 3 array of microscope images to observe a larger area and choose the region to scan.



