Application Note | High-Temperature Profilometry

In Situ High Temperature Surface Morphology Using 3D Profilometry

Watching Solder Melt, Wet and Spread in 3D from Room Temperature to 400 °C

Line sensor scanning a heated solder sample

Research & Experimental Testing

Duanjie Li

Visual Design & Editorial

Andrew Shore

Introduction

The 3D dimensional shape of a material or device is vital to its service quality and functionality. Surface roughness and texture can change at elevated temperatures. For example, shape-memory alloys (also called smart metals) remember their original shape and return to the pre-deformed shape when heated. Most polymeric materials melt and degrade rapidly at high temperatures, leading to change of shape and structure. Therefore, it is crucial to investigate the mechanism of shape deformation and the limit of the materials exposed to high temperature environments so as to select the proper material for the target applications.

Importance of 3D Non-Contact Profilometry for Surface Morphology at High Temperatures

High temperature environments can change the surface texture, roughness and shapes of materials, resulting in device malfunction and mechanical failures. To ensure the quality of materials or devices used at elevated temperatures, accurate and reliable in situ monitoring of the shape evolution at high temperatures is needed to provide insight into the mechanism of material deformation. Moreover, real-time monitoring of surface morphologies at high temperatures is very useful in materials processing, such as laser machining.

The NANOVEA 3D Non-Contact Profilometers measure the surface morphology of materials without touching the sample, avoiding introducing additional scratches or shape alteration which may be caused by contact technologies such as a sliding stylus. Their capacity of non-contact measurement also makes it possible to measure the shape of melted samples.

ℹ️ Learn more about surface profiling lab services.

Measurement Objective

In this application, the NANOVEA ST400 non-contact profilometer equipped with a line optical sensor is used to monitor the surface evolution of a Sn60Pb40 solder sample heated from room temperature up to 400 °C. We showcase the capacity of the NANOVEA non-contact profilometer in providing fast and precise 3D profile measurement at elevated temperatures.

NANOVEA ST400 Standard

Optical 3D Profilometer

NANOVEA ST400 modular optical profilometer for non-contact 3D surface measurement
Optical line sensor projecting a line of light onto a solder sample on a heating stage
Fig. 1: Optical line sensor scanning on the surface of the solder sample.

Results and Discussion

A Sn60Pb40 solder sample was heated up from room temperature to 400 °C. The evolution of its 3D topography was monitored in situ using the NANOVEA ST400 non-contact profilometer equipped with a line sensor. The optical line sensor generates a bright line of 192 points, as shown in Fig. 1. These 192 points scan the sample surface at the same time, leading to significantly increased scan speed. An 8×8 mm² area scan at each temperature in this study can be finished in 5 s.

The 3D view of the solder surface topography at representative temperatures is shown in Fig. 2, providing an ideal tool for users to observe the solder sample from different angles. The 2D cross section profiles are compared in Fig. 3. The solder maintains its shape as heated up from room temperature until ~180 °C, where its melting point is reached. The solder rapidly changes its shape and adheres to the copper substrate underneath it, forming a smooth spherical cap. As the temperature continues to rise to ~300 °C, rosin flux starts to burn at the edge of the solder, leading to roughening of the surface. When a temperature of ~400 °C is reached, the solder spreads to the surrounding area and lowers its height. Meanwhile, the rosin flux on and around the solder continues to burn, leaving behind traces of graphite. The evidence of such observations is displayed in Fig. 4.

Four 3D views of the solder at room temperature, 180, 300 and 400 °C, changing from a coil to a smooth cap to a spread puddle
Fig. 2: Evolution of the solder sample at (a) room temperature, (b) 180 °C, (c) 300 °C and (d) 400 °C.
Cross-section profiles of the solder at the four temperatures
Fig. 3: 2D profile of the solder sample at (a) room temperature, (b) 180 °C, (c) 300 °C and (d) 400 °C.
Photos of the solder at 300 °C with burnt flux at its edge and at 400 °C spread out with graphite traces
Fig. 4: The solder sample at (a) 300 °C and (b) 400 °C.

Conclusion

In this application, we have showcased that the NANOVEA ST400 3D Non-Contact Profilometer monitors the evolution of the surface morphology on a solder sample heated from room temperature to 400 °C. The optical line sensor generates a line consisting of 192 light spots that scan the sample surface at the same time. It significantly increases the scan speed, enabling monitoring of the 3D shape evolution of the solder in situ. It is observed that the solder melts at ~180 °C. Its shape continues to change as the temperature progressively increases to 300 °C and 400 °C.

The data shown here represents only a portion of the calculations available in the analysis software. NANOVEA Profilometers measure virtually any surface in fields including Semiconductor, Microelectronics, Solar, Fiber Optics, Automotive, Aerospace, Metallurgy, Machining, Coatings, Pharmaceutical, Biomedical, Environmental and many others.

Frequently Asked Questions About High-Temperature Surface Measurement

Why measure surface shape at high temperature?

Heat changes texture, roughness and shape: shape-memory alloys recover their form and polymers melt or degrade. In-situ monitoring reveals how and when materials deform, which matters for material selection and for processes such as laser machining.

How can a molten sample be measured?

A non-contact profilometer measures without touching, so it adds no scratches or deformation and can even measure the shape of melted samples.

How fast is each scan?

The line sensor projects 192 points at once, so an 8 × 8 mm² area scan took 5 s at each temperature, fast enough to follow the solder as it changed.

What happened to the solder as it heated?

It kept its shape until about 180 °C, then melted and wetted the copper beneath into a smooth spherical cap; around 300 °C the rosin flux began to burn and roughen the edge, and by about 400 °C the solder spread out and lowered.

What was the sample?

A Sn60Pb40 solder on a copper substrate, heated from room temperature to 400 °C.

Which instrument was used?

The NANOVEA ST400 3D non-contact profilometer with an optical line sensor.